Abstract
The Kelvin Probe Method for
measuring the difference in work function between two surfaces, known as
the "Contact Potential Difference" (CPD), is a well-known technique in
surface science. It permits accurate and precise measurements of CPDs in
a wide range of technologically important environments, including air,
(non-conducting) liquids, and vacuum.
Unfortunately, the
traditional implementation of the method, using a lock-in amplifier, suffers
from two significant sources of systematic error: the spacing dependence
of the feedback gain involved in a lock-in technique [1],
and the spacing dependence introduced by stray capacitances, or "micro
phonics" [2].
Any Kelvin Probe measurements which do not account for and eliminate these
effects, may contain essentially unpredictable systematic errors of several
hundreds of millivolts.
We describe an implementation
of the Kelvin Probe technique which relies on off-null measurements and
which permits essentially spacing-independent measurements to be made.
The Kelvin Probe Technique
Figure 1
shows a schematic diagram of the Kelvin Probe apparatus. A probe is held
above the surface of interest and oscillated mechanically. When dissimilar
materials are used, a contact potential difference (CPD) due to the difference
in work functions exists between the probe and sample, and an ac current
will flow through the circuit. If an external potential VB
is applied to the probe that exactly cancels out the CPD, the current will
be zero. VB is then
exactly equal to the CPD.
Figure 1. Schematic diagram
of the Kelvin Probe
Traditionally, a lock-in
amplifier is used to measure the signal and negative feedback used to null
out the signal. Then, simply by measuring the feedback voltage and dividing
by the gain of the preamp, the CPD could be determined. There are several
problems associated with this technique.
-
This method suffers from a poor
signal to noise ratio due to having to work at the null point where, by
definition, the signal to noise ratio is zero.
-
Two sources of systematic error
related to the mean spacing between the plates exist.
The first characteristic of
the lock-in technique is that the capacitance of the probe/sample interface
can significantly affect the gain of the feedback loop.
[1] Reducing this dependence requires using
higher preamp gains, causing greater instability in the feedback loop.
Another dependence on mean spacing results from unintended vibrations
in the head. These vibrations induce their own currents resulting in a
false signal. These "stray capacitances" can lead to errors in the measured
CPD of the order of tens or hundreds of millivolts[2].
The Off-Null Variation
We have implemented a variation
on the traditional Kelvin Probe method based on an "off-null" technique
[2].
The amplitude of the current measured by the preamp is proportional to
the potential difference across the plates. By measuring the signal amplitude
as a function of applied backing potential and interpolating to the point
of zero current, the backing potential required to null out the signal
is found. This, as before, is simply the CPD.
The off-null method
has the advantage that no feedback is used, and thus the gain of the preamp
is independent of the mean spacing of the probe and sample. In addition,
for off-null conditions, the signal to noise ratio is very good and, therefore,
interpolation can be performed very accurately.
Even this method suffers,
however, from systematic errors due to stray capacitances. In Figure
2, the line labeled Unoptimized Probe shows the CPD between
a stationary copper substrate and an oscillating stainless steel probe,
measured as a function of the mean spacing. The probe and sample were clean,
but were covered by an uncharacterized oxide layer. As can be seen from
the data, the measured CPD varies by almost one hundred millivolts over
a range of spacings of approximately 250 microns. Clearly, for any accurate
measurement, this uncertainty is unacceptable.
Figure 2.
Kelvin Probe Output
Minimizing the Effect of Stray
Capacitances.
The first step in minimizing
stray capacitances is to reduce the stray vibrations themselves. All loose
wires should be rigidly held in place, with care being taken not to add
unnecessary capacitance to the circuit (for instance, glueing wires to
a surface can add a huge capacitance, making the RC time constant so large
that the system is essentially unusable). In addition, the wire that connects
the moving parts to the rigid head, should be made short, with minimal
excess length. It should also be very thin, and kept far from any stationary
objects to reduce the capacitance to those objects. By making it short,
the number of resonant modes and frequencies is reduced; by making it thin,
the resonant frequencies can be raised above the range in which the probe
will operate.
Even in a well-designed
head, vibrations will exist and introduce stray capacitances. Therefore,
we have developed a technique which can reduce the effect of the vibrations
even further. First, the false signal due to stray capacitance is measured
by withdrawing the sample far from the probe so that the true KP signal
is negligible. With stray capacitances present, there will still be a "signal".
The amplitude of this signal is measured as a function of frequency over
the range of frequencies available - in this case from 50 to 500 Hz. The
minima in this spectrum show the frequencies at which the stray capacitances
contribute the least. By taking data at these non-resonant frequencies,
the effects of stray capacitance on the measured CPD can be reduced even
further.
Figure 3. Two different
Kelvin Probes
Figure
3 shows a graph of the stray signal amplitude for two different heads,
as a function of the probe oscillation frequency, at constant probe driving
voltage. For a constant driving voltage, the amplitude of oscillation of
the probe will vary with frequency. In particular, at the resonant frequency,
the amplitude will be large.
The line labeled Unoptimized
Probe shows the signal for a head containing several loose wires, capable
of vibrating relative to a grounded metal object. The line labeled Optimized
Probe shows the stray capacitance signal for the same head after optimization.
Two important features are apparent.
-
For the first probe, the average
signal level is significantly higher than the second.
-
Secondly, the optimized head
still shows a large peak in amplitude near the resonance of the probe.
We believe this is due to the very large (>2mm) amplitude of vibration
of the probe at this frequency.
The effect of the stray capacitance
signal is to introduce an error into the measured CPD:
Equation 1
where h
is the mean spacing between probe and sample and
is a constant dependent on the amplitudes of oscillation, the relative
"square areas" of the probe and vibrating wire, and the mean spacing of
the vibrating wire from the nearby surface. This curve is also displayed
in Figure 3. In this fit there are two free fit
variables,
and
Stray.
For the line shown in Figure 3,
=
0.005 and
Stray=
-0.615).
Conclusions:
The dependence of the measured
contact potential difference (CPD) on the mean spacing between the probe
and sample represents a major source of systematic error in the Kelvin
Probe technique. This error is essentially unpredictable and may be as
large as hundreds of millivolts. As a result, any measurements made which
do not explicitly account for these errors may contain errors of this magnitude.
There are two main
sources of this dependence: the dependence of the feedback amplifier gain
on the capacitance of the probe/sample gap for lock-in techniques; and
the effects of stray capacitances (micro phonics) on the measured CPD.
We show how the use of an
"off-null" technique eliminates the former, while careful head design and
selection of probe oscillation frequency can eliminate the latter. As a
result, we obtain CPD measurements that are independent to within a few
millivolts, of the mean spacing over a typical experimental range of 250
microns.
References
1. Rossi,
F. Contact potential measurement: Spacing dependence errors. Rev
Sci Instrum 63 (9), 4174-4181 (1992).
2. Baikie,
I. D. et al, Analysis of stray capacitance in the Kelvin method.
Rev Sci Instrum 62 (3), 725-735 (1991)